Recently, GoGo Instruments successfully completed an important delivery—providing customized radiative tensile heating stages and customized radiative heating stages for a university's Plasma-Surface Interaction under Extreme Conditions (PSIEC) facility for first-wall materials. This batch of equipment will be integrated into the PSIEC system to support research on irradiation damage and fuel retention behavior of plasma-facing materials (PFM) under coupled particle/stress/thermal loads in multi-field environments.
◑ Customer Requirements: Simulating Extreme Fusion Reactor Conditions and Breaking Through Bottlenecks in Multi-Field Coupled In-Situ Testing
In the actual operational scenario of the PSIEC facility, the user faces the following core challenges:
◈ Difficulty in synergistic high-temperature and stress loading: The PSIEC facility is designed to apply pre-stress loads to specimens through the sample stage, and to simulate the high-temperature first-wall conditions of a fusion reactor via active heating. Conventional temperature control devices struggle to simultaneously maintain 800°C high temperatures and precisely apply 1500 N tensile forces under vacuum and magnetic field environments.
◈ High compatibility requirements for magnetic field and vacuum environments: Testing must be conducted under vacuum at 2×10⁻⁵ Pa and a magnetic field of 800–1200 Gauss, with samples maintained at 800°C and exposed to plasma beams (power 1–5 kW). Conventional heating and mechanical testing equipment are difficult to operate stably under strong magnetic fields and high vacuum, and are prone to electromagnetic interference.
◈ Difficulty in weak signal acquisition and precise control: Plasma irradiation experiments demand extremely high temperature stability (±0.1°C) and displacement accuracy (±1 μm), making it difficult for conventional equipment to achieve high-precision temperature control and mechanical feedback under such complex operating conditions.
◑ Solution: GoGo Instruments Customized Radiative Tensile Heating Stage and Radiative Heating Stage
GoGo Instruments custom-developed two core equipment units for a university's PSIEC facility, enabling high-temperature tensile mechanical testing and high-temperature irradiation testing functions respectively on single platforms. The key parameters are as follows:
• Customized Radiative Tensile Heating Stage (H800/F1500N/V4-C01)
Heating method: Resistive heating
Temperature control range: RT ~ 800°C
Temperature stability: ±0.1°C
Temperature sensor: Thermocouple
Tensile force range: 0 ~ 1500 N
Tensile force accuracy: 0.5% F.S.
Displacement travel: 10 mm
Tensile speed: 0.1 ~ 10 mm/min
Displacement accuracy: ±1 μm
• Customized Radiative Heating Stage (H800/V4-C01)
Heating method: Resistive heating
Temperature control range: RT ~ 800°C
Temperature stability: ±0.1°C
Temperature sensor: Thermocouple
Optical path: Transmissive optical path, with four Φ9.5 mm holes reserved on the platinum alloy protective plate

Both devices are compatible with the following stringent testing environments:
• Vacuum level: 2×10⁻⁵Pa
• Magnetic field strength: 800–1200 Gauss
• Sample holding temperature: 800℃
• Plasma beam exposure power: 1–5 kW
Advantage 1: High-Temperature, High-Precision Temperature Control Ensuring Reliability of Plasma Irradiation Experiments
Both devices employ resistive heating, with a temperature control range covering room temperature to 800°C, temperature stability of ±0.1°C, and temperature resolution of 0.1°C. Combined with a PID precision temperature control algorithm and thermocouple sensors, they enable long-term stable temperature control under vacuum and magnetic field environments, meeting the stringent temperature control requirements of the PSIEC facility for high-temperature irradiation experiments on first-wall materials.
Advantage 2: Integrated Tensile Mechanical Module Enabling Synergistic Particle/Stress/Thermal Loading
The customized radiative tensile heating stage integrates a tensile force range of 0–1500 N, tensile force accuracy of 0.5% F.S., a displacement travel of 10 mm, an adjustable tensile speed of 0.1–10 mm/min, and displacement accuracy of ±1 μm. This device can simultaneously apply pre-stress loads to specimens during plasma irradiation, and together with the active heating function of the sample stage, it realizes the synergistic particle/stress/thermal loading experimental environment designed for the PSIEC facility.
Advantage 3: Deep Adaptation to Vacuum and Magnetic Field Environments, Ensuring Data Purity Under Complex Operating Conditions
Both devices are specifically designed for the PSIEC facility's high vacuum of 2×10⁻⁵ Pa and magnetic field of 800–1200 Gauss, ensuring stability and reliability of temperature control and mechanical feedback during plasma beam exposure (power 1–5 kW). The radiative heating stage is additionally equipped with a transmissive optical path and four Φ9.5 mm holes on the platinum protective plate, reserving interfaces for future optical diagnostics and further expanding the facility's in-situ characterization capabilities.
◈ Delivery Site: Installation, Commissioning, and Operation Training
After completing the installation and commissioning of the equipment, the GoGo Instruments technical team conducted specialized training on the operation procedures, temperature control program settings, tensile control parameter adjustments, and equipment usage specifications under vacuum and magnetic field environments for the radiative tensile heating stage and the radiative heating stage. This ensures that the research team of the university's PSIEC facility can quickly get up to speed and independently conduct plasma irradiation experiments on first-wall materials under multi-field coupled conditions.
The two custom-built devices delivered this time will be formally incorporated into the in-situ testing system of the university's PSIEC facility, supporting its frontier research in areas such as irradiation damage of plasma-facing materials, fuel retention behavior, and service performance of fusion reactor wall materials.
● On-Site Equipment Integration